How much turbidity can the Nitratax sc Sensor handle in the sample process?

Document ID

Document ID TE2364

Version

Version 4.0

Status

Status Published

Published Date

Published Date 06/08/2018
Question
How much turbidity can the Nitratax sc Sensor handle in the sample process?
Summary
Nitratax sc and turbidity
Answer
The Nitratax sc Sensor can handle turbidity up to 1000 NTU or FNU and still read reliably across all measurement pathlengths. The actual maximum turbidity value cannot be stated as it will depend on the sludge structure.  Because the Nitratax sc sensor has two channels (measure and reference) and is working with wavelengths that are very close together, the sensor is able to see, allocate and compensate for turbidity up to 1000 NTU or FNU.
 