Question
How do I sample using the ultrasonic sensor during a high alarm period with the AS950 sampler?
Summary
AS950 and ultrasonic sensor sampling
Answer
For programming the sampler:
1 Start a datalogger program for the ultrasonic so it is recording the readings.
2 Add an alarm for the ultrasonic that is the reading that will trigger the sampler (high flow rate for starting sampler)
3 In the sampler program, go to "Pacing" then choose "Event". In the Event selection choose the Ultrasonic alarm that you made.
Once this is done the sampler will start when the "alarm" (ultrasonic) is triggered.