What can cause the Astro TOC/1950+ to read zero and what can fix this?

Document ID

Document ID TE2311

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/06/2019
Question
What can cause the Astro TOC/1950+ to read zero and what can fix this?
Summary
Astro TOC 1950+ Reading Zero
Answer
Zero readings can be due to a calibration issue, or due to component failure.
1. Check the calibration. If the calibration was performed while there was a leak in the system it is possible the curve is a flat line. This will cause the analyzer to read zero, and this is especially true if you are able to perform a gas calibration.
2. The live value is not based on calibration, it is the raw value. Do a liquid cal to see if it will calibrate and read the standard properly. What ppm does it show? If 0, there may be an issue with the bench. Recalibrate it.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.