Can the special abrasive paper be ordered separately for the CL10 chlorine sensor maintenance?

Document ID

Document ID TE2188

Published Date

Published Date 01/04/2022
Question
Can the special abrasive paper be ordered separately for the CL10 chlorine sensor maintenance?
Summary
CL10sc chlorine sensor maintenance
Answer
When sensor readings are unstable or the calibration slope is too low, it is time to service the chlorine sensor by polishing the electrode with the special abrasive paper. This special abrasive paper is not sold as an individual item. It is included as a component of the membrane kit. It is also included with the purchase of a new sensor. 
  
New sensor kits: Catalog Number/ Order Code 9150400 - sensor, free chlorine CLF10 sc and Catalog Number/ Order Code 9150300 - sensor, total chlorine, CLT10 sc.
CL10 replacement membrane kit part numbers are listed in the below articles
What membrane cap should be used with the CL10 electrode?
What is included in the CL10 membrane replacement kits?

 