How do you change the units on the APA6000 analyzer from mg/L to ug/L?

Document ID

Document ID TE2186

Version

Version 5.0

Status

Status Published

Published Date

Published Date 03/13/2018
Question
How do you change the units on the APA6000 analyzer from mg/L to ug/L?
Summary
Display and output resolution options on the APA6000 analyzer.
Answer
In the recorder setup menu, choose the appropriate output being used. That screen allows you to change your minimum output value, arrow over to the decimal point and then you can arrow up to give it a value (such as zero). Do the same with the maximum output value.

For example: for the APA 6000 Low Range Hardness analyzer - If the desired reading is 50 ug/L to 10,000 ug/L instead of .05-10 mg/L, the decimal point  can be removed by adjusting over to the decimal point, and then pressing the up arrow to give a value. To change resolution from X.XXX to XXX.X, then the decimal position can be changed with the same method.