Can formazin standards be used to calibrate for TSS?

Document ID

Document ID TE2154

Published Date

Published Date 10/10/2019
Print
Question
Can formazin standards be used to calibrate for TSS?
Summary
TSS sc and Suspended Solids Calibration
Answer
No. The formazin standards are used for turbidity calibration only. Suspended solids calibration must be done using a gravimetric analysis in order to be accurate. The gravimetric analysis is a lab method that determines the exact amount of suspended solids in the process water. Methods 8158 and 8164 are gravimetric methods that are US EPA accepted.