Question
What types of interferences will affect my DO reading using an LDO2 probe?
Summary
LDO2sc sensor and interferences
Answer
No interferences from the following: H2S, pH, K+, Na+, Mg2+, Ca2+,NH4 +, Al3+, Pb2+, Cd2+, Zn2+, Cr (total), Fe2+, Fe3+, Mn2+, Cu2+, Ni2+, Co2+, CN–, NO3–, SO4 2–, S2–, PO4 3–, Cl–, Anion Active Tensides, Crude Oils, Cl 2 < 4 ppm. Other than these, anything that would dissolve or destroy the plastic sensor cap itself, would destroy the sensor.