Service Reminders are displayed when counters expire on 5500sc analyzer. How can I reset after service is done?

Document ID

Document ID TE2143

Version

Version 5.0

Status

Status Published

Published Date

Published Date 05/16/2018
Question
Service Reminders are displayed when counters expire on 5500sc analyzer. How can I reset after service is done?
Summary
5500sc Service Reminder reset.
Answer
When counters expire, reminders will display.
To reset the counters, do the following:

-Hit DIAG key
-Move down until SERVICE is highlighted and hit ENTER
-Highlight SERVICE PART and hit ENTER
-Highlight the reminder you want to reset and hit ENTER
-It will show a screen with last and next service, days remaining and RESET SERVICE DATE?
-At Bottom of screen it will have choice of NO and YES. Check in front of YES and hit ENTER

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.