What do you multiply the offset by when performing a Solitax sc sensor turbidity calibration?

Document ID

Document ID TE2133

Version

Version 5.0

Status

Status Published

Published Date

Published Date 04/11/2019
Question
What do you multiply the offset by when performing a Solitax sc sensor turbidity calibration?
Summary
Solitax sc Turbidity Calibration
Answer
Multiply the offset by negative one (-1). This value is missing from the current User Manual. The reason for this is that when you put the probe into the DI water your goal is to make the probe read zero. If the probe is reading 5 NTU in DI water you will make the offset -5, and the probe will then read zero in the DI water.