Question
What TOC standard should be used to calibrate the BioTector TOC analyzer?
Summary
BioTector TOC standards recommendation.
Answer
You want to use a TOC standard that is about 80% of the range of the BioTector analyzer. The standard can either be prepared according to the directions listed in the user manual, or purchased from Hach. Please contact technical support with the range of your analyzer so that we can provide a Product # for the appropriate TOC standard.