Question
What does 'inhibit' do on AS950 sampler?
Summary
AS950 sampler using 'inhibit' feature contacts for external start/stop.
Answer
The 'inhibit' function is a short circuit between pins 2 and 4 on the auxiliary port that allows the user to run the sample program when an external switch is closed and suspend the program when the switch is open. The program will resume from where it left off when the switch is closed again. The 'inhibit' function can be used to start a sampler program, to trigger samples, or to start and stop a sampler program depending on the programming settings and users needs