What should the color of the reagents for the CL17 chlorine analyzer be?

Document ID

Document ID TE2045

Published Date

Published Date 04/11/2019
Print
Question
What should the color of the reagents for the CL17 chlorine analyzer be?
Summary
CL17 reagent color information
Answer
The buffer solution should be clear and colorless. The indicator solution, after the DPD powder is added, can vary in color from clear to a light pink. The color will vary from lot to lot, but this does not indicate a quality problem, and it will not affect your chlorine results. If the indicator solution turns a dark purple then the indicator solution has been contaminated and should be replaced with a fresh bottle.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.