Is it possible to add analog outputs to an an older style CL17 chlorine analyzer that goes through a signal input module (SIM) and AquaTrend system?

Document ID

Document ID TE2040

Version

Version 5.0

Status

Status Published

Published Date

Published Date 03/07/2019
Question
Is it possible to add analog outputs to an an older style CL17 chlorine analyzer that goes through a signal input module (SIM) and AquaTrend system?
Summary
CL17 additional outputs for SIM and AquaTrend system.
Answer
Additional equipment can be added to the output loop in series as long as the total impedance of the system does not exceed 500 Ohms. If the impedance is over 500 Ohms, an additional signal output module (SOM) must be added to the network, or use a powered third party 4 - 20 mA signal splitter.