The CL17 chlorine analyzer is reading correctly, but the pumps are not turning on, how can this be resolved?

Document ID

Document ID TE2038

Version

Version 4.0

Status

Status Published

Published Date

Published Date 03/07/2019
Question
The CL17 chlorine analyzer is reading correctly, but the pumps are not turning on, how can this be resolved?
Summary
CL17 Not Operating Pumps
Answer
Check to make sure the relays are working correctly by (pressing 918 DIAG on the Older CL17 analyzer or Test Relays on the newer CL17 analyzer). Then check the set points and make sure that the alarm is set to LOW and not HIGH in the Relay set up.