Why is my pH sensor slow to respond during calibration?

Document ID

Document ID TE1944

Published Date

Published Date 03/06/2019
Print
Question
Why is my pH sensor slow to respond during calibration?
Summary
Differential or Combination pH Sensor Operation
Answer
When walking through a calibration of a pH sensor and it is very slow to respond, the normal procedure would be to clean the sensor tip, and perhaps change the salt bridge and cell solution on a differential sensor. If this does not make a difference, go through the configuration menu and check to see what is programmed for FILTER. If it has a value, that is the signal average time. Changing the value to zero will speed up sensor response.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.