Why is my pH sensor slow to respond during calibration?

Document ID

Document ID TE1944

Published Date

Published Date 01/05/2022
Question
Why is my pH sensor slow to respond during calibration?
Summary
Differential or Combination pH Sensor Operation
Answer
When walking through a calibration of a pH sensor and it is very slow to respond, the normal procedure would be to clean the sensor tip, and perhaps change the salt bridge and cell solution on a differential sensor. If this does not make a difference, go through the configuration menu and check to see what is programmed for FILTER. If it has a value, that is the signal average time. Changing the value to zero will speed up sensor response.

See the following articles and links for more information on pH sensor maintenance
Is there a video on how to change a salt bridge for the pHD sensors?
pH Prove How to:  Improving Performance via Routine cleanings

Was this answer helpful?