How do I add the auto flush option to an AF7000 that was originally ordered without the option?

Document ID

Document ID TE1926

Version

Version 5.0

Status

Status Published

Published Date

Published Date 02/27/2018
Question
How do I add the auto flush option to an AF7000 that was originally ordered without the option?
Summary
AF7000 Analyzer and Auto Flush Option
Answer
Order Product # 25061030 - AUTOFLUSH CARD, 110V S VALVE, AF7000.
Perform the following steps to add from the Main Screen:
1. Press Menu/Cancel.
2. Press the down button until READ is displayed on the bottom display. Press Enter.
3. Press the down button until DATA is displayed on the bottom display. Press Enter.
4. Press the down button until OPT1 is displayed on the bottom display. Press Enter.
5. Use the up or down arrows to add the appropriate number to the value displayed on the screen. As an example, an AF7000 with Auto Flush, 4-20 mA Output card, and 2 alarm relays would display 16 on the screen.
6. Press Menu/Cancel until the instrument returns to the main screen.
7. Cycle the power on the instrument. The options for the newly added card should show up in the menu structure (see user manual for full structure).
 