How many measurements will the 5500sc or 9610sc analyzer log in memory?

Document ID

Document ID TE1904

Published Date

Published Date 01/06/2022
Question
How many measurements will the 5500sc or 9610sc analyzer log in memory?
Summary
5500sc silica/phosphate analyzer and memory log
Answer
The data storage capacity of the 5500/961X silica/phosphate analyzers is 128k bytes for data logs and 64k bytes for event logs. This will be approximately 12,800 measurements, or about 80 days of measurements when using a standard continuous measurement cycle.