Does the PhotonMaster™ require routine servicing or calibration?

Document ID

Document ID TE1885

Published Date

Published Date 08/20/2019
Print
Question
Does the PhotonMaster™ require routine servicing or calibration?
Summary
PhotonMaster™ routine service and calibration
Answer
The PhotonMaster™does not require routine service. That being said, it is recommended to clean the PhotonMaster™ and verify linearity an annual basis using the LuminUltra Luminometer Maintenance Kit (Product # LMK-5).

Instructions for cleaning the PhotonMaster™ are given in the following article:

How should the LuminUltra Photonmaster™ be cleaned?

Calibration of the instrument is done by the user with the UltraCheck™ 1 standard. Perform one calibration per day or for each set of samples analyzed at the same time.

 