What is the level of sensitivity of the LuminUltra 2nd Generation ATP tests?

Document ID

Document ID TE1798

Version

Version 5.0

Status

Status Published

Published Date

Published Date 08/20/2019
Question
What is the level of sensitivity of the LuminUltra 2nd Generation ATP tests?
Summary
Sensitivity of LuminUltra 2 nd Generation ATP tests.
Answer
The sensitivity of the LuminUltra 2 nd Generation ATP tests will vary depending on the test kit being used:
  • Quench-Gone™ Aqueous (QGA): The QGA test kit has a sensitivity of 0.1 pg/mL. This is the base sensitivity; higher sensitivity can be achieved through additional filtration.
  • Quench-Gone™ Organic Modified (QGO-M): The QGO-M test kit has a sensitivity of 0.25 pg/mL.
  • QuenchGone21™ Specialty (QG21S): The QG21S test kit has a sensitivity of 20 pg/mL.
  • QuenchGone21™ Industrial (QG21I): The QG21I test kit has a sensitivity of 20 pg/mL.
  • QuenchGone21™ Wastewater (QG21W): The QG21W test kit has a sensitivity of 0.22 ng tATP/mL and 2.02 ng dATP/mL.
  • Deposit &Surface Analysis (DSA™): The DSA test kit has a sensitivity of 100 pg tATP/unit.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.