What is the expected precision and sensitivity of the Total Kjeldahl Nitrogen Method 8075?

Document ID

Document ID TE1716

Version

Version 4.0

Status

Status Published

Published Date

Published Date 06/13/2019
Question
What is the expected precision and sensitivity of the Total Kjeldahl Nitrogen Method 8075?
Summary
Total Kjeldahl Nitrogen Method 8075 Performance
Answer
The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users may get different results under different test conditions. For program 399 with a 76 mg/L NH3-N standard, the precision (95% confidence interval) was 70 mg/L NH3-N. The sensitivity concentration change per 0.010 Abs change was 1 mg/L NH3-N.
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.