How is the System Verification Chemkey® used with the SL1000?

Document ID

Document ID TE1507

Published Date

Published Date 12/20/2021
Question
How is the System Verification Chemkey® used with the SL1000?
Summary
Using the Verification Chemkey® to validate the SL1000 Portable Parallel Analyzer™ (PPA)
Answer
To use the System verification Chemkey (Product # 9436800) follow the instructions below.
1. Push the Menu button
2. Select Verify Cal
3. Insert the Verification Chemkey® in slot 1 and allow instrument to verify slot.
4. Repeat step 3 for slots 2-4.

If this procedure is not followed and the Verification Chemkey® is inserted while the meter is in measurement mode, the meter will display an error message, "Verify Only".