Can extreme pH effect the color formation of the PhosVer® 3 phosphate reagent?

Document ID

Document ID TE1474

Version

Version 5.0

Status

Status Published

Published Date

Published Date 08/09/2018
Question
Can extreme pH effect the color formation of the PhosVer® 3 phosphate reagent?
Summary
Effect of extreme pH on PhosVer® 3 phosphate reagent
Answer
Extreme pH can effect the color formation of the PhosVer® 3 phosphate reagent in methods 8048 and 8190. The sample should be near pH 7 prior to adding the PhosVer® 3 phosphate reagent. If samples have been acidified from digestion for total phosphorus determination, 5.0 N sodium hydroxide should be used to adjust to pH 7. After the addition of the PhosVer® 3 phosphate reagent, the sample should be pH 1-2.