What does the error "Wrong CDM PROM version" upon starting the TIM8xx or TIM9xx titrator mean?

Document ID

Document ID TE1400

Published Date

Published Date 07/26/2018
Print
Question
What does the error "Wrong CDM PROM version" upon starting the TIM8xx or TIM9xx titrator mean?
Summary
Wrong CDM PROM version error on Radiometer titrator
Answer
This error typically occurs with older titrators where the titrator's software was just updated. This occurs because a component of the main board is not capable of accepting the new software because it is outdated. This component needs to be replaced with a more current version of the hardware. This requires the instrument to be sent into our Service department for repair. Please contact Technical Support at 800.227.4224 ext 6122 for further assistance.