Does the BW680 Ballast Water Handheld Fluorometer measure phaeopigments?

Document ID

Document ID TE1335

Version

Version 3.0

Status

Status Published

Published Date

Published Date 09/11/2018
Question
Does the BW680 Ballast Water Handheld Fluorometer measure phaeopigments?
Summary
Phaeopigment measurement with the BW860
Answer
The BW680 (Product # 2037700) measures variable fluorescence derived from active chlorophyll a, found in physiologically competent phytoplankton. Degradation products of chlorophyll a, such as phaeopigments, will not trigger a response in the BWI parameter. The BWI parameter can be likened to the variable fluorescence term, FV, which is the difference between Fm and Fo, the max and min values of variable fluorescence, respectively (Fv=Fm-Fo).