How can a Pocket Colorimeter™ II be reset to factory default?

Document ID

Document ID TE1245

Version

Version 6.0

Status

Status Published

Published Date

Published Date 09/27/2018
Question
How can a Pocket Colorimeter™ II be reset to factory default?
Summary
Reset Pocket Colorimeter™ II to factory default
Answer
To reset a Pocket colorimeter II, press and hold the menu button until USER then CAL displays on the screen. Press the Blue zero button twice until dFL is displayed. Press the read or green button to accept this reset. The unit will go back to the reading mode once this reset has been performed.
Not all colorimeters have the ability to be reset see: Can all Pocket Colorimeter II instruments be reset to factory default?

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.