Question
How can a Pocket Colorimeter™ II be reset to factory default?
Summary
Reset Pocket Colorimeter™ II to factory default
Answer
To reset a Pocket colorimeter II, press and hold the menu button until USER then CAL displays on the screen. Press the Blue zero button twice until dFL is displayed. Press the read or green button to accept this reset. The unit will go back to the reading mode once this reset has been performed.