What is the icon on the right-hand side of the Pocket Colorimeter™ II screen that either looks like a graph or an arrow pointing to the lower left?

Document ID

Document ID TE1243

Published Date

Published Date 09/20/2018
Print
Question
What is the icon on the right-hand side of the Pocket Colorimeter™ II screen that either looks like a graph or an arrow pointing to the lower left?
Summary
Pocket Colorimeter™ II has icon that looks like a graph or an arrow
Answer
The graph icon on the right portion of the screen of the Pocket Colorimeter™ II either indicates that a user entered calibration has been applied or a standard calibration adjust has been applied.
If an adjustment is not expected, it is recommended to rest the unit to factory defaults, Article Link, and remove the standard calibration adjustment, Article Link