In the cyanide method 8027 will sample temperature impact measurement?

Document ID

Document ID TE1097

Published Date

Published Date 09/09/2020
Print
Question
In the cyanide method 8027 will sample temperature impact measurement?
Summary
Sample temperature for cyanide testing
Answer
Method 8027 states to use a water bath to keep the temperature for the reaction at the optimum 25°C for best results. Samples less than 23°­­C need longer reaction times, and samples more than 25°C give low results.