Why did the instrument verification using Test Filter Set for the DR3900 and DR6000 Spectrophotometers fail?

Document ID

Document ID TE1073

Published Date

Published Date 12/20/2021
Question
Why did the instrument verification using Test Filter Set for the DR3900 and DR6000 Spectrophotometers fail?
Summary
Instrument fails verification when using Test Filter Set
Answer
When using the Test Filter Set (Product # LZV537) in a spectrophotometer, it is necessary to enter in the nominal values of the filter set before running the checks. Using the default nominal values or a previous lot nominal values will result in incorrect values. Enter the nominal values in the menu option and repeat the filters.

 