All of the data is gone from the DR900, how did this happen?

Document ID

Document ID TE975

Published Date

Published Date 08/09/2018
Question
All of the data is gone from the DR900, how did this happen?
Summary
Data missing from the DR900
Answer
The DR900 firmware versions prior to 1.060 had a defect that caused the stored programs to overwrite and corrupt SPM memory space, which resulted in the loss of data records. The only other way to delete stored records is to restore the DR900 to factory settings, which deletes data, favorites, and user programs.