How are the Glass Fiber Filters, Preweighed for Standard Methods TSS Procedure prepared?

Document ID

Document ID TE930

Version

Version 3.0

Status

Status Published

Published Date

Published Date 10/11/2018
Question
How are the Glass Fiber Filters, Preweighed for Standard Methods TSS Procedure prepared?
Summary
Preweighed filter paper preparation
Answer
Each Pre-Weighed Total Suspended Solids (TSS) Filter consists of 1.5 μm binder-free, glass microfiber filters which have been vacuum rinsed three times with 20 mL aliquots of reagent grade water, dried at 105°C for at least one hour, desiccated to room temperature, weighed to the nearest 0.1 mg and then placed in an aluminum pan with an affixed, heat resistant bar code label preprinted with filter ID number and the filter weight.

All washing, drying, desiccating, and weighing are done in accordance with Standard Methods 2540 D and the filters are ready to use out of the box. The pretreatment of these filters cuts labor costs and TSS analysis time in half.

Each box of pre-weighed filters will contain a Certificate of Conformity indicating strict adherence to Standard Methods 2540 D specifications.

The Glass Fiber Filters, Preweighed for Standard Methods TSS procedure, are available in the following diameters and pack sizes:
 
Product # Diameter (mm) Filters/Pack
2546110 42.5 100
2546100 47 100
2546120 55 100
2546130 76 100
2546140 90 100
2546150 110 100
2546210 42.5 400
2546200 47 400
2546220 55 400
2546230 76 400
2546240 90 400
2546250 110 400


Please note: some of the items listed in this article are obsolete.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.