What is the procedure for the SpecCheck Gel Secondary Standard Set, Fluoride?

Document ID

Document ID TE859

Version

Version 4.0

Status

Status Published

Published Date

Published Date 07/24/2018
Question
What is the procedure for the SpecCheck Gel Secondary Standard Set, Fluoride?
Summary
Procedure for SpecCheck Gel Secondary Standard Set, Fluoride
Answer
Below is the procedure for SpecCheck Gel Secondary Standard Set, Fluoride (Product # 2712500). Please note that this procedure is different than some of the other SpecCheck procedures.

1. Place STD 1 into the cell holder with the alignment mark facing the keypad, cover the cell with the instrument cap.
2. Press ZERO. The display will show "0.00" or "0.0" depending on the range.
3. Place the blank cell into the cell holder. Cover the cell with the instrument cap.
4. Press READ/ENTER. Record the concentration measurement.
5. Repeat steps 1-4 with cells labeled STD 2 and STD 3.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.