Does the SL1000 Portable Parallel Analyzer™ (PPA) need to be calibrated by the user?

Document ID

Document ID TE831

Version

Version 4.0

Status

Status Published

Published Date

Published Date 07/24/2018
Question
Does the SL1000 Portable Parallel Analyzer™ (PPA) need to be calibrated by the user?
Summary
SL1000 Portable Parallel Analyzer™ (PPA) Calibration
Answer
The instrument comes fully calibrated from the factory. We have built the calibration curves from scratch and have tested over 450,000 Chemkey® Reagents to build these curves.  A customer does have the ability to adjust the curve; however we discourage doing so, as a 4-8 point curve adjustment is never going to be as accurate as a 100,000 point curve.
 
Probes that are connected to the SL1000 Portable Parallel Analyzer™ (PPA) will need to be calibrated just as they would be if attached to an HQD meter.