If I use the silica Series 5500 analyzer molybdate 3 reagent, how do I account for the reagent blank?

Document ID

Document ID TE800

Version

Version 3.0

Status

Status Published

Published Date

Published Date 09/06/2018
Question
If I use the silica Series 5500 analyzer molybdate 3 reagent, how do I account for the reagent blank?
Summary
Reagent blank using Series 5500 molybdate reagent
Answer
The ULR Silica program 645 (method 8282) allows for a reagent blank value to be entered into the spectrophotometer. The reagent blank is based on the reagent part number. See below for reagent blank options (as provided in the “before starting the test” section of the procedure). *Verify the accuracy of the ULR method using a 500ug/L standard solution. The Silica Reagent 3, Molydate, 2.9L used for the series 5000 Silica Analyzer can be used for method 8282 for more information review the following article: Is there an ultra low range (ULR) silica method that utilizes the Silica Analyzer series 5000 or Series 5500 analyzer reagent set?

S5000 Silica Reagent 3, Molybdate, 2.9L (Product # 199503) = multiply reagent blank on label by 1.09
Molybdate 3 Reagent, 100mL (Product # 199532) = use the blank value on the bottle
Molybdate 3 Reagent, 1L (Product # 199553) = use the blank value on the bottle

 