Does Hach have a general statement that explains how the solids calibration correlates to measured value?

Document ID

Document ID TE752

Published Date

Published Date 09/13/2018
Print
Question
Does Hach have a general statement that explains how the solids calibration correlates to measured value?
Summary
TSS portable solids measurement explanation
Answer
Yes, a single point calibration draws a straight line from the probes zero point to the point of calibration; however the mathematical transformation of the different signals of the different detectors is nonlinear. When a calibration curve consists of 3 points the curve is created by connecting the straight lines. The course may not hit the zero e.g. because the medium itself is colored and absorbs light.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.