Question
Is there any precaution to switching between TSS portable curves while still taking a continuous measurement?
Summary
Switching between TSS portable curves during continuous measurement
Answer
When switching curves during a measurement, the next measurement is done with the new calibration curve. The respective data set is stored respectively. Data sets consist of Calibration curve name/Meas. Value/date and time. If an individual measurement site at the plant is assigned to one calibration curve, data may easily be identified. If you have two measurement points using the same calibration curve, you may distinguish the measurement sites only by the time the measurement was taken. In the field this should work fine since measurements are done in a predefined order.