What is the TSS portable system reset code?

Document ID

Document ID TE735

Published Date

Published Date 10/09/2019
Print
Question
What is the TSS portable system reset code?
Summary
TSS portable system reset code
Answer
The code to reset the TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Level System Product # LXV322.99.00002 is 4711.

To use the system reset code, the following steps must be performed:
  1. Select MENU and confirm with ENTER/ON..
  2. Use the up and down arrow buttons to highlight SYSTEM and confirm with ENTER/ON..
  3. Use the up, down, left, and right arrow keys to enter in the system reset code 4711.
  4. Once the system reset code has been entered, press ENTER/ON to confirm.
  5. The meter will then display M-Type. Press ENTER/ON to select this option.
  6. The meter will then display TSS as well as several other dashed options. Confirm  the TSS option pressing ENTER/ON.
  7. The meter will then power off. Press ENTER/ON button to power the meter on.
  8. The meter will be reset to German. To change the language to English, select MENU and confirm with ENTER/ON.
  9. Use the up and down arrow buttons to highlight SPRACHE/LANGUAGE and then confirm with ENTER/ON.
  10. Select ENGL./ENGLISH and confirm with ENTER/ON to restore the meter to English.