How should the Pocket Pro and Pro+ meters be stored when not in use?

Document ID

Document ID TE696

Published Date

Published Date 10/09/2019
Print
Question
How should the Pocket Pro and Pro+ meters be stored when not in use?
Summary
Pocket Pro and Pro+ meter storage
Answer
The pH models (including the Multi 1/2) and ORP models should be stored with a couple drops of tap water or deionized water in the cap to keep the bulb hydrated.

pH Electrode Storage Solution, 500 mL (Product # 2756549) is not recommended to use with the pocket pro testers. This is because the storage solution is 3M KCl and the pocket pro testers have a 4M KCl electrolyte. Storing in lower concentration storage solution will cause a mV offset decrease the life of the tester.

Temperature, conductivity and TDS  models can be stored dry.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.