Why does the DR 6000 software require a 10 mm square cell on Copper program 135, but the Hach procedure is requiring a 1 inch sample cell?

Document ID

Document ID TE662

Version

Version 3.0

Status

Status Published

Published Date

Published Date 04/24/2018
Question
Why does the DR 6000 software require a 10 mm square cell on Copper program 135, but the Hach procedure is requiring a 1 inch sample cell?
Summary
DR 6000 copper program 135 requires a 10 mm square sample cell
Answer
Older software versions had this message in error. The 1 inch cell is correct for this program. Updating the software version will eliminate this error.