In the cadmium reduction method what causes low nitrate results on samples containing salt?

Document ID

Document ID TE518

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/30/2018
Question
In the cadmium reduction method what causes low nitrate results on samples containing salt?
Summary
Cadmium reduction for nitrate in seawater samples
Answer
Chloride in the sample at levels greater than 100 mg/L slows the reaction with the NitraVer5 and NitraVer6 reagents. The best solution is to calibrate the instrument using nitrate standards that have been spiked with chloride to the same level as the samples, or switch to a different analysis method. Some nitrate methods include details how to create a calibration curve with the chloride matrix.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.