What is the difference between standard adjust and user calibration?

Document ID

Document ID TE502

Version

Version 5.0

Status

Status Published

Published Date

Published Date 02/27/2018
Question
What is the difference between standard adjust and user calibration?
Summary
Standard adjust and user calibation
Answer
Standard adjust maintains the shape of the factory calibration curve, but the whole curve is adjusted slightly so that the instrument reads the exact value of a standard solution. Typical limits for calibration curve adjustment are 10-20 percent. These limits ensure that if a standard were prepared incorrectly, the instrument would reject the adjustment. User calibration puts an entirely new calibration curve into the instrument, and does not have rejection limits if standards are prepared incorrectly. User calibration is not recommended for this reason, however it may be required by regulations.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.