What can be done if the pH reading on a Platinum Series electrode is unstable and erratic?

Document ID

Document ID TE478

Version

Version 3.0

Status

Status Published

Published Date

Published Date 08/28/2018
Question
What can be done if the pH reading on a Platinum Series electrode is unstable and erratic?
Summary
Platinum Series electrode pH reading is unstable and erratic
Answer
Most likely the electrolyte gel in the Platinum Series Electrode is not making contact with the sample or buffer, or there is an air bubble in the gel. To correct this, click the dispenser on the electrode until the electrolyte gel comes out and the pH reading is stable. See the electrode instruction manual for more information on the electrolyte gel cartridge.

 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.