How often should gel be dispensed from a Platinum Series electrode?

Document ID

Document ID TE477

Version

Version 4.0

Status

Status Published

Published Date

Published Date 08/28/2018
Question
How often should gel be dispensed from a Platinum Series electrode?
Summary
Dispensing gel from a Platinum Series electrode
Answer
Click the dispenser two or three times every day before calibration, thereafter it is possible to measure several samples without clicking the dispenser unless the electrode is rinsed vigorously between samples. Click the dispenser every five to ten samples after calibration.