What is the correct way to clean the sensor on a dissolved oxygen LDO® probe?

Document ID

Document ID TE421

Published Date

Published Date 07/11/2022
Question
What is the correct way to clean the sensor on a dissolved oxygen LDO® probe?
Summary
Cleaning the sensor on a dissolved oxygen LDO® probe
Answer
Wipe the probe with a soft, wet cloth to remove deposits. For stubborn deposits, use a dilute detergent solution, rinse well, and soak in tap water for approximately 10 minutes. Do not use organic solvents or abrasives because these will destroy the sensor.

Alcohol can be used to clean the probe as long as the probe is not submerged for more than 2 minutes and cleaning is not more than 2 times per day. This information can be found in a UK ENG Disinfection of LDO electrode_DOC042.52.20136 along with a UK ENG Poster: Troubleshooting + Maintenance of LDO and Clark Probes_DOC142.52.20160that can be found on the Hach website.