Can Stablcal® be used instead of Gelex to to perform verification?

Document ID

Document ID TE409

Version

Version 6.0

Status

Status Published

Published Date

Published Date 10/18/2018
Question
Can Stablcal® be used instead of Gelex to to perform verification?
Summary
Turbidimeter verification with Stablcal® or Gelex standards
Answer
Gelex standards, developed well before Stablcal® standards became available, were designed to provide a convenient way to check the calibration of turbidimeters on a daily or weekly basis. At that time, the only other way to check turbidimeter calibration was to prepare formazin standards from dilutions of a 4000 NTU standard. Now that stabilized formazin standards are available as Stablcal® in sealed ampules, they also eliminate the need for dilution and can also be used for checking the calibration of a turbidimeter on a daily or weekly basis. Stablcal® standards must first be inverted before use according to directions, however, to ensure that the formazin polymer is suspended in solution.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.