Is it really necessary to use organic-free water for the total organic carbon (TOC) blank vial? I have good quality deionized water that I would prefer to use.

Document ID

Document ID TE394

Version

Version 3.0

Status

Status Published

Published Date

Published Date 10/02/2018
Question
Is it really necessary to use organic-free water for the total organic carbon (TOC) blank vial? I have good quality deionized water that I would prefer to use.
Summary
Organic-free water for Total Organic Carbon blank
Answer
Organic-free water must be used for the blank vial in the TOC tests using methods 10128, 10129 and 10173 ; deionized water can have low concentrations of organic matter that will cause the blank to read high and cause low test results. Not using organic-free water is one of the most common problems that people have with this test. Organic-free water should be packaged in glass bottles; water packaged in plastic containers will show contamination and cause low results.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.