Can the PhosVer® 3 Phosphate Reagent Powder Pillowsbe used with the pour-thru cell or Sip10?

Document ID

Document ID TE383

Version

Version 5.0

Status

Status Published

Published Date

Published Date 08/09/2018
Question
Can the PhosVer® 3 Phosphate Reagent Powder Pillowsbe used with the pour-thru cell or Sip10?
Summary
Using PhosVer® 3 powder pillows with the pour-thru cell or Sip10
Answer
The PhosVer® 3 Phosphate Reagent Powder Pillows chemistry is not listed in the pour-thru manual or Sip10 manual as “Not Suitable”, however, the PhosVer® 3 chemistry may not fully dissolve and is not recommended. Undissolved reagent may be a source of contamination for future samples or can interfere with the readings in pour-thru cell.

Please also see: Which chemistries are incompatible with the SIP10?


 