Which iron method is the correct one to use - FerroVer®, FerroZine®, or TPTZ?

Document ID

Document ID TE313

Version

Version 7.0

Status

Status Published

Published Date

Published Date 08/16/2018
Question
Which iron method is the correct one to use - FerroVer®, FerroZine®, or TPTZ?
Summary
Choosing the correct iron method
Answer
The correct iron test for your sample type will depend on a few factors.

-The FerroVer®, FerroZine®, and TPTZ methods for iron differ primarily in sensitivity. FerroZine® is the most sensitive, then TPTZ, and then FerroVer®.
-FerroVer® is the only USEPA-accepted method for reporting total iron (note: the sample must be digested, and a colorimeter or spectrophotometer must be used when reporting).
-Both the FerroVer® and TPTZ methods react with iron that is sequestered with complexing agents such as EDTA, but FerroZine®will not.
-The FerroVer® method is the simplest to use and reacts with all but the most very resistant forms of iron in the sample.
-The FerroZine® method uses a liquid reagent that has a strong sulfur smell.

All of the iron methods can be found in the Water Analysis Handbook.
Attachments

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.