Is there an extra step in the Winkler titration if the floc in the dissolved oxygen sample will not settle?

Document ID

Document ID TE274

Published Date

Published Date 09/20/2018
Print
Question
Is there an extra step in the Winkler titration if the floc in the dissolved oxygen sample will not settle?
Summary
Floc in Winkler dissolved oxygen sample will not settle
Answer
The floc that forms after adding the first two reagents usually settles to the bottom of the bottle after sitting undisturbed for a few minutes. In samples of high density, however, (for example seawater or very cold samples) the floc may not settle. If the floc does not settle after 5 minutes, proceed to the next step, adding sulfamic acid. Be careful that none of the floc sticks to the stopper when it is removed from the bottle.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.